Open original FDA record ↗Recall record
- FDA identifier
- Z-1521-2015
- Company in FDA record
- Ortho-Clinical Diagnostics
- FDA classification date
- Apr 25, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Software Anomaly; It is possible for the device to process samples with a cartridge other than the intended cartridge, potentially leading to erroneous patient results. If this anomaly occurred previously, an indication would be a series of results that were similar across multiple samples (i.e., results could be believable or outside of the Reportable Range for the intended assay).
Source facts
- Action
- On 3/10/2015, URGENT PRODUCT CORRECTION NOTIFICATION letters (Ref. CL2015-047, dated 3/10/2015) were sent to users notifying them of the issue and advising them of the required actions. Customers are to complete and return the Confirmation of Receipt no later than March 23, 2015. The firm is currently working on a resolution and will issue a follow up notification upon the software's availability. For questions, please contact Customer Technical Services at 1-800-421-3311.
- Initiated
- 2015-03-10
- Root cause
- Software in the Use Environment
- Terminated
- 2018-06-01
- Distribution
- Worldwide Distribution -- USA, Puerto Rico, Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Mexico, Republic of Panama, Singapore, Spain, and Venezuela.
- Recall number
- Z-1521-2015
- Recall status
- Terminated
- CDRH record ID
- 134489
- FDA classified
- 2015-04-25
- Recall event number
- 70685