Z-1522-2021: CareLink SmartSync Device Manager (24970A). used to interrogate cardiac implantable electronic devices (

Recall record Z-1522-2021 for Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF). View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1522-2021
Company in FDA record
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
FDA record creation date
Apr 30, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a SmartSync software issue that results in an overestimation in the displayed longevity of these devices during an approximate 6-month window of time before the device triggers its Recommended Replacement Time (RRT).

Source facts

Action
Beginning 06-APR-2021, Medtronic notified risk managers and practice managers of the field corrective action notification via 2-day UPS delivery of the consignee letter. Confirmation delivery of the consignee letter will be obtained. Medtronic Field Representatives have been instructed to begin installation of the software update. Outside the United States (OUS), communication began on 06- APR-2021 using locally approved methods. Medtronic Field Representatives will begin installation of the software update as it is available locally. Consignees were advised that Medtronic representatives can assist in installing the software update on programmers. During device interrogation with an updated programmer at implant or the next regularly scheduled clinic visit, a patients device will automatically receive the updated software. A courtesy notification will be provided to physicians. No confirmation will be required for the courtesy notification.
Initiated
2021-04-06
Root cause
Software design
Distribution
Worldwide
Recall number
Z-1522-2021
Recall status
Open, Classified
CDRH record ID
186464
FDA record created
2021-04-30
Recall event number
87654

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗