Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
FDA record creation date
Apr 30, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II
Reason stated in the source
There is a SmartSync software issue that results in an overestimation in the displayed longevity of these devices during an approximate 6-month window of time before the device triggers its Recommended Replacement Time (RRT).
Source facts
Action
Beginning 06-APR-2021, Medtronic notified risk managers and practice managers of the field corrective action notification via 2-day UPS delivery of the consignee letter.
Confirmation delivery of the consignee letter will be obtained. Medtronic Field Representatives have been instructed to begin installation of the software update.
Outside the United States (OUS), communication began on 06- APR-2021 using locally approved methods. Medtronic Field Representatives will begin installation of the software update as it is available locally.
Consignees were advised that Medtronic representatives can assist in installing the software update on programmers. During device interrogation with an updated programmer at implant or the next regularly scheduled clinic visit, a patients device will automatically receive the updated software.
A courtesy notification will be provided to physicians. No confirmation will be required for the courtesy notification.
Initiated
2021-04-06
Root cause
Software design
Distribution
Worldwide
Recall number
Z-1522-2021
Recall status
Open, Classified
CDRH record ID
186464
FDA record created
2021-04-30
Recall event number
87654