Z-1522-2022: INTERA 3000 Hepatic Artery Infusion Pump, 30 mL, Sterile, Rx Only. implantable infusion pump.

Recall record Z-1522-2022 for Intera Oncology, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1522-2022
Company in FDA record
Intera Oncology, Inc.
FDA classification date
Aug 22, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Higher than expected flow rate.

Source facts

Action
Customers were notified by a combination of letter, call, and/or email of this recall on July 10, 2022. Customers were informed via notification that Intera 3000 Hepatic Artery Infusion Pumps are being recalled due to complaints reporting higher-than-expected flow rates. Customers are asked to secure affected pumps in inventory to ensure they are not implanted and to contact Intera Oncology at 1-800-660-2660 to arrange return of product. Intera requests physicians continue to monitor the pump flow rate per standard refill procedure, and continue to monitor for liver toxicity to adjust dosing as per standard protocols.
Initiated
2022-07-10
Root cause
Process design
Terminated
2023-09-12
Distribution
US distribution to states of: AL, CA, IL, KY, MN, MO, NC, NJ, NY, OH, OR, PA, & TX.
Recall number
Z-1522-2022
Recall status
Terminated
CDRH record ID
195297
FDA classified
2022-08-22
Recall event number
90642

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗