Skip to content

Public FDA record

Z-1523-2013: AMICUS Exchange Kit; Product Usage: This kit is designed for use with the AMICUS separator for the Ther

Recall record Z-1523-2013 for Fenwal Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1523-2013
Company in FDA record
Fenwal Inc
FDA classification date
Jun 8, 2013
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Fenwal has initiated a voluntary Urgent Product Recall for lot FA12C07070 of product code R4R2339, Amicus Exchange Kits. Fenwal identified a labeling issue with this batch of Product Code R4R2339 (which was cleared for European use only) in which this batch was distributed within the United States market. The problem is identified as labeling that did not include the written description for the s

Source facts

Action
Fenwal sent an Urgent Product Recall letter dated May 24, 2013 to all affected customers. The letter identified the affected product, problem and actions to be taken. Customers were instructed to immediately check inventory, discontinue the use of this prod.uct, return the affected product to Fenwal, and complete and return the attached reply form. For questions call 1-800-333-6925.
Initiated
2013-05-24
Root cause
Process change control
Terminated
2016-08-01
Distribution
USA Nationwide Distribution
Recall number
Z-1523-2013
Recall status
Terminated
CDRH record ID
118494
FDA classified
2013-06-08
Recall event number
65296

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.