Skip to content

Public FDA record

Z-1523-2014: Datascope System 98/98xt, CS 100/CS100i and CS 300 Intra-Aortic Balloon Pump. The Intra-Aortic Balloo

Recall record Z-1523-2014 for Maquet Datascope Corp - Cardiac Assist Division. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1523-2014
Company in FDA record
Maquet Datascope Corp - Cardiac Assist Division
FDA classification date
May 21, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Potential mechanical failure of the fan assembly associated with the power supply. A fan assembly failure could result in the power supply overheating and cause the IABP to shut down without warning.

Source facts

Action
Maquet Inc. sent an Urgent Medical Device Field Correction letter dated March 21, 2014, to all affected customers. The letter identified the affected product, problem, and actions to be taken.. Customers were instructed that their Service Representatiave would contact them to schedule the replacement of the fan assembly at no cost to the customer. Upon completion customers were requested to sign a service repair order to verify satisfactory completion of the work. Customers with questions within Puerto Rico, were instructed to contact Technical Support in Mexico at +52-55-9000-8970, ext 104. For questions regarding this recall call 973-709-7652. A field correction letter was sent out on April 15, 2014 to additional consignees that did not receive the March 15, 2014 letter. Maquet issued press on May 9, 2014.
Initiated
2014-03-21
Root cause
Nonconforming Material/Component
Terminated
2018-03-20
Distribution
Worldwide Distribution - USA (nationwide) and approximately 104 countries.
Recall number
Z-1523-2014
Recall status
Terminated
CDRH record ID
126390
FDA classified
2014-05-21
Recall event number
67815

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗