Potential mechanical failure of the fan assembly associated with the power supply. A fan assembly failure could result in the power supply overheating and cause the IABP to shut down without warning.
Source facts
Action
Maquet Inc. sent an Urgent Medical Device Field Correction letter dated March 21, 2014, to all affected customers. The letter identified the affected product, problem, and actions to be taken.. Customers were instructed that their Service Representatiave would contact them to schedule the replacement of the fan assembly at no cost to the customer. Upon completion customers were requested to sign a service repair order to verify satisfactory completion of the work. Customers with questions within Puerto Rico, were instructed to contact Technical Support in Mexico at +52-55-9000-8970, ext 104.
For questions regarding this recall call 973-709-7652.
A field correction letter was sent out on April 15, 2014 to additional consignees that did not receive the March 15, 2014 letter.
Maquet issued press on May 9, 2014.
Initiated
2014-03-21
Root cause
Nonconforming Material/Component
Terminated
2018-03-20
Distribution
Worldwide Distribution - USA (nationwide) and approximately 104 countries.
Recall number
Z-1523-2014
Recall status
Terminated
CDRH record ID
126390
FDA classified
2014-05-21
Recall event number
67815