Skip to content

Public FDA record

Z-1523-2022: Power electronic box, Part #AS-0066, subassembly for the MyoCycle MC-2.

Recall record Z-1523-2022 for Myolyn Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1523-2022
Company in FDA record
Myolyn Inc.
FDA record creation date
Aug 16, 2022
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Screws mounting an internal power supply may come loose during shipping resulting in increased risk of high voltage electrical shock.

Source facts

Action
The recalling firm issued letters dated 7/19/2022 via email on 7/19/2022 informing the consignee of the reason for recall, risk to health, and action to be taken by the customer/user. The letter informs the addressee that the firm will ship a replacement power electronics box to all customers affected by the recall and the customers are asked to replace their original power electronics box with the new box and then return the original box to the recalling firm. The letter explains that instructions for replacing the box and a return UPS shipping label are enclosed with the replacement power electronics box. For customers who are unable or unwilling to replace their original power electronics box, a MYOLYN representative will travel to the customer location to perform the replacement. For customers who have received a device that has exhibited any signs of failure, they were instructed to discontinue use and disconnect it from the main electronical supply. If no power failures have been observed, customers can continue to use their device but should be cautious when transporting it to prevent the power supply mounting screws from loosening. The customer is to acknowledge receipt and understanding of the letter by clicking on the link in the letter or scanning the QR code to open the Recall Acknowledgement Form.
Initiated
2022-07-19
Root cause
Device Design
Terminated
2024-09-23
Distribution
Distribution was made to AZ, CA, CT, FL, GA, ID, IL, KY, MA, MI, MN, MO, MN, NC, NJ, NM, NY, OH, OR, TX, UT, VA, WA, and WI. There was no foreign/military/government distribution.
Recall number
Z-1523-2022
Recall status
Terminated
CDRH record ID
195135
FDA record created
2022-08-16
Recall event number
90620

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗