Z-1524-2020: Medtronic CareLink 2090 Programmer. Used to interrogate and program Medtronic and Vitatron implantable d

Recall record Z-1524-2020 for Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF). View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1524-2020
Company in FDA record
Medtronic Inc., Cardiac Rhythm and Heart Failure (CRHF)
FDA record creation date
Mar 19, 2020
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Medtronic Conexus Telemetry has been determined to contain two primary cyber vulnerabilities: improper access control and the cleartext transmission of sensitive information.

Source facts

Action
Medtronic initially notified customers on about 03/21/2019 via a "Security Bulletin" regarding vulnerabilities to the proprietary Medtronic Conexus radio frequency wireless telemetry protocols associated with some Medtronic ICDs (implantable cardioverter defibrillators) and CRT-Ds (cardiac resynchronization therapy defibrillators). On about 01/30/2020, Medtronic notified customers with affected programmers of the field corrective action notification via 2-day UPS delivery of the consignee letter. On 21 March 2019, Medtronic posted a Security Bulletin regarding the vulnerabilities. Customers were instructed that Medtronic Representatives can assist in installing the Conexus Telemetry update on programmers Once the programmer has been updated, the updated software version can be identified on the programmer by selecting the programmer icon on the programmer screen. Then, select software from the dropdown menu. During device interrogation with an updated programmer at implant or the next regularly scheduled clinic visit, a patients device will automatically receive the updated software. Additionally the recall notice should be shared with all those who need to be aware within your organization or with any organization where programmers have been transferred. Questions regarding this information should be directed to the local Medtronic Representative or Medtronic Technical Services at 800-638-1991.
Initiated
2020-01-30
Root cause
Software design
Distribution
Worldwide distribution. US Nationwide including Puerto Rico, Argentina, Brazil, Colombia, Trinidad and Tobago, Venezuela, Dominican Republic, Bahamas, Peru, Bermuda, Barbados, Ecuador, Cayman Islands, Curacao, Guatemala, Honduras, Panama, El Salvador, Costa Rica, Bolivia, Chile, Uruguay, Paraguay, Mexico, Jamaica, Japan, Belgium, Greece, Germany, Switzerland, United Kingdom, Spain, Italy, Luxembourg, Poland, Austria, Suriname, Netherlands, Aruba, Hungary, Liechtenstein, Sweden, Czech Republic, Ireland, France, French Polynesia, New Caledonia, Guadeloupe, Reunion, Mauritius, Kenya, South Africa, Uganda, Namibia, Finland, Norway, Portugal, Iceland, Denmark, Faroe Islands, Greenland, Macedonia, Iraq, Pakistan, Sudan, Jordan, Morocco, Belarus, Lebanon, Sri Lanka, Kazakhstan, Bulgaria, Egypt, Turkey, Libya, Serbia, Tunisia, Cyprus, Oman, Slovenia, Algeria, Malta, Croatia, Saudi Arabia, Syria, Kuwait, Latvia, Ukraine, Andorra, Burkina Faso, Nepal, Mayotte, French Guiana, Martinique, Bosnia And Herzegovina, Qatar, United Arab Emirates, Georgia, Yemen, Slovakia, Iran, Canada, Lithuania, United States, Cote D'Ivoire, Nicaragua, Zimbabwe, Azerbaijan, Albania, Ghana, India, Israel, Guam, American Samoa, Korea, Armenia, Mozambique, Nigeria, Bangladesh, Australia, Taiwan, Cambodia, Hong Kong, Malaysia, Viet Nam, Singapore, New Zealand, Indonesia, Philippines, Angola, Rwanda, Northern Mariana Islands, Senegal, Cameroon, Botswana,Russian Federation, Montenegro, China, Turkmenistan, St. Maarten, Tanzania, Kyrgyzstan, Netherlands Antilles, Cuba, Romania, Mongolia, Kosovo, Tajikistan, Benin, Ethiopia, Moldova, Afghanistan, Democratic Republic of Congo, and Thailand
Recall number
Z-1524-2020
Recall status
Open, Classified
CDRH record ID
179694
FDA record created
2020-03-19
Recall event number
84944

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗