Open original FDA record ↗Recall record
- FDA identifier
- Z-1524-2022
- Company in FDA record
- Luminex Corporation
- FDA record creation date
- Aug 17, 2022
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
SARS-CoV-2 assay that detects SARS-CoV-2 ORF1ab and N genes may not detect the ORF1ab gene. A positive results if either SARS-CoV-2 genes is detected, but If both genes are mistakenly not detected, a false negative result would be generated.
Source facts
- Action
- On 7/14/22 recall notices were emailed to consignees who were asked to review any results from the affected lot to confirm the validity of the results. In addition, all remaining cassettes from this lot should be discarded. Customers with questions are encouraged to contact the firm at: 1-877-785-2323 (U.S. and Canada), +1-512-381-4397 (Outside U.S. and Canada), support@luminexcorp.com
- Initiated
- 2022-07-14
- Root cause
- Nonconforming Material/Component
- Distribution
- U.S. Nationwide distribution in the states of NE, TX, MS, TN, KY, ME, CA.
- Recall number
- Z-1524-2022
- Recall status
- Open, Classified
- CDRH record ID
- 195445
- FDA record created
- 2022-08-17
- Recall event number
- 90698