Open original FDA record ↗Recall record
- FDA identifier
- Z-1525-2010
- Company in FDA record
- Bard Access Systems
- FDA classification date
- Apr 29, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
8F PowerPort stems were mixed with 6F PowerPorts and packaged.
Source facts
- Action
- The firm, Bard Access Systems, Inc. (BAS) sent a "URGENT: Powerport isp M.R.I with 6F Chronoflex Catheter RECALL" letter dated March 18, 2010, to the customers. The letter describes the product, problem and actions to be taken by customers. The customers were instructed to immediately examine their inventory and quarantine all products subject to this recall and notify any customers they may have distributed the product to; DO NOT USE THIS PRODUCT OR FURTHER DISTRIBUTE ANY OF THE IMPLICATED LOTS; remove and return the affected lots to BAS, and complete and Fax the Reply Form and Inventory Reconciliation Form, if unable to Fax call and report information. Only unused products, in its original packaging, should be returned to BAS. Should you have any questions or require assistance in this matter, please contact BAS Customer Service by calling our toll-free number 1-800-290-1689.
- Initiated
- 2010-03-18
- Root cause
- Mixed-up of materials/components
- Terminated
- 2010-10-01
- Distribution
- Nationwide distribution: AZ, CA, DC, FL, MA, MD, NC, NE, NV, NY, PA and SC.
- Recall number
- Z-1525-2010
- Recall status
- Terminated
- CDRH record ID
- 90222
- FDA classified
- 2010-04-29
- Recall event number
- 55175