Reason stated in the source
Neonatal Flow Sensors supplied for use with the Neonatal Enhancement for the Engstrm Carestation may result in the sensors providing invalid values and alarm messages when exposed to high flow rates.
Source facts
- Action
- Consignees were sent a Medical Device Correction Notification letter on July 2, 2007. The firm's short term recommendation is that Neonatal Flow Sensors may continue to be used with added viligance around potential patient circuit disconnections and the associated alarms. Ther firm recommends that the MV exp high and Circuit leak alarms be set appropriately for each patient in order to detect a patient disconnect. The long term recommendation was to order the replacement part when available and dispose of the old part when the replacement is received.
- Initiated
- 2007-07-02
- Root cause
- Nonconforming Material/Component
- Terminated
- 2012-07-03