Cursor on the dialysis screen may disappear and not respond to user input
Source facts
Action
Fresenius Medical Care contacted customers by mailing a Customer Notification letter "Urgent correction FMNCA Model 2008T Hemodialysis Machine" via Certified-Mail, Return Receipt requested on January 19, 2011. The Technical Services department began contacting consignees by telephone on January 13, 2011, to schedule an appointment for installing the electronic board on the 2008T machines at the customers' facilities.
For questions call Technical services 1-800-227-2572.
Initiated
2011-01-19
Root cause
Device Design
Terminated
2013-07-08
Distribution
Worldwide Distribution - USA and Canada
Recall number
Z-1526-2011
Recall status
Terminated
CDRH record ID
97578
FDA classified
2011-03-04
Recall event number
57851