Z-1526-2014: Medtronic, External neurostimulator, ENS (37021) & ENS (37022). Including the external neurostimulators

Recall record Z-1526-2014 for Medtronic Neuromodulation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1526-2014
Company in FDA record
Medtronic Neuromodulation
FDA classification date
Apr 29, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Medtronic neurostimulators (used with pain and deep brain stimulation therapy) have a cycling feature that, when enabled, allows stimulation output to be toggled on and off based on the programmed settings. The use of this feature is determined by the physician and set using the clinician programmer. Current labeling indicates the use of cycling improves device longevity and recharge interval

Source facts

Action
Medtronic sent a Medical Device Correction letter dated February 2014 to all affected consignees. The letter identified the product, the problem, and the action to be taken by the consignee. Consignees were instructed to do the following: If you have patients with cycling enabled or wish to use cycling, use the updated calculations and graphs available online at professional.medtronic.com/cycling to assess the impact of cycling on device longevity. Medtronic will update the product labeling in the second half of 2014. " Remind your patients to continue to check their battery status. Instructions on how to check battery status can be found in the patient programmer or recharger system manual. " Pain Stimulation Trialing Only: The cycling feature should not be enabled on the External Neurostimulator (ENS) when estimating device longevity because the resulting longevity estimate for non-rechargeable devices may not provide accurate information. Consignees were also instructed to acknowledge receipt of the notice by completing the attached reply form and returning it to Medtronic using the contact details on the reply form. For questions consignees can contact Medtronic Neuromodulation Technical services at 1-800-707-0933 weekdays 7am - 6pm CST. For questions regarding this recall call 800-707-0933.
Initiated
2014-02-27
Root cause
Labeling False and Misleading
Terminated
2014-11-07
Distribution
Nationwide Distribution - all states including DC and Puerto Rico.
Recall number
Z-1526-2014
Recall status
Terminated
CDRH record ID
126207
FDA classified
2014-04-29
Recall event number
66841

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗