Open original FDA record ↗Recall record
- FDA identifier
- Z-1527-2010
- Company in FDA record
- Ge Healthcare, Llc
- FDA classification date
- May 11, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Unanticipated shut-down: A specific lot of GE Healthcare Aisys and Avance machines control board wiring harness have a defect, which can cause the machine to unexpectedly shut down, terminating ventilation, anesthetic delivery, and potentially patient monitoring.
Source facts
- Action
- Consignees were sent a GE Healthcare "Urgent Medical Device Correction" letter dated March 12, 2010. The letter was addressed to Chief of Anesthesia, Health Care Administrator/Risk Manager, Director of Biomedical/Clinical Engineering. The letter listed the Safety Issue, Affected Product Details, Safety Instructions, Product Correction and Contact Information.
- Initiated
- 2010-03-12
- Root cause
- Process design
- Terminated
- 2012-01-30
- Distribution
- Worldwide distribution: USA (CA, FL, TN), CANADA, DENMARK, FINLAND, FRANCE, ITALY, SAUDI ARABIA, SOUTH AFRICA, SPAIN, UNITED ARAB EMIRATES, and UNITED KINGDOM.
- Recall number
- Z-1527-2010
- Recall status
- Terminated
- CDRH record ID
- 90654
- FDA classified
- 2010-05-11
- Recall event number
- 55355