Open original FDA record ↗Recall record
- FDA identifier
- Z-1527-2014
- Company in FDA record
- Laurimed LLC
- FDA classification date
- Apr 30, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
The reciprocation mechanism in the PolypVac Microdebrider may turn off prematurely due to a component failure in the device handle.
Source facts
- Action
- All direct accounts were notified on April 8, 2014 with an Urgent: Medical Device Recall of PolypVac Microdebrider letter, sent by first class mail. The letter identified the affected product, the reason for the recall, and stated that the firm is in the process of implementing corrections. The firm is requesting that all unused lots be returned to them. Replacement devices will sent to customers. In addition, customers are to complete the attached Lot Information sheet. Questions should be directed to bdubois@laurimed.com or 650-587-5296.
- Initiated
- 2014-04-07
- Root cause
- Component design/selection
- Terminated
- 2014-05-14
- Distribution
- Distributed in the states of CA, VT, NC, DO, TX, FL, PA, OR, AZ, IL, ID, and VA.
- Recall number
- Z-1527-2014
- Recall status
- Terminated
- CDRH record ID
- 126647
- FDA classified
- 2014-04-30
- Recall event number
- 67950