Skip to content

Public FDA record

Z-1528-2018: EEA" Hemorrhoid and Prolapse Stapler Set with DST Series" Technology 4.8MM, Product Number HEM3348, GTIN

Recall record Z-1528-2018 for Covidien LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1528-2018
Company in FDA record
Covidien LLC
FDA record creation date
Apr 26, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a potential for improper welding of the yellow staple guide to the instrument. Use of a device with an improperly welded staple guide may result in improper staple formation potentially leading to bleeding or anastomotic leak.

Source facts

Action
Notification letters were distributed on 4/9/18. The letters instructed customers to: Medtronic requests that you quarantine and return any unused products of the item codes and lots detailed above. Unused products from the affected item codes and lots should be returned as described in the Required Actions section below. If you have distributed Covidien EEA" Hemorrhoid and Prolapse Stapler Set with DST Series" Technology listed above, please promptly forward the information from this letter to those recipients. All unused products from the affected item codes and lots must be returned. Required Actions: 1. Please quarantine and discontinue use of the affected item code and lots listed on page one. 2. Please return affected product as indicated below. 3. Complete the Recalled Product Return Form even if you do not have inventory. Purchased directly from Medtronic: Complete the Recalled Product Return Form and fax to: 800-895-6140 Ship affected product with RGA# provided by Customer Service to: Medtronic Attn: Field Returns 195 McDermott Road North Haven, CT 06473 USA Complete form and check the box indicating no inventory Fax to 800-895-6140 or email to feedback.customerservice@Covidien.com Exception: Customers with zero inventory, fax to 203-492-7719 or email to FCAMITG@Covidien.com Purchased from a distributor: Complete all fields on the form and contact your distributor directly to arrange for return of product Complete form and check the box indicating no inventory Fax to 203-492-7719 or email FCAMITG@Covidien.com
Initiated
2018-04-09
Root cause
Nonconforming Material/Component
Terminated
2020-05-12
Distribution
Worldwide Distribution - US Nationwide in the US states of : AL, CA, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, NC, NE, NJ, NY, OH, OR, SC, TN, TX, UT, and VA. The devices were distributed to the following foreign countries: UAE, Australia, Austria, Belgium, Bulgaria, Canada, Czech Republic, Denmark, Estonia, Finland, France, Germany, Indonesia, Italy, Korea, Malaysia, Netherlands, Poland, Portugal, Russia, Saudi Arabia, Senegal, Singapore, Slovakia, South Africa, Spain, Sweden, Switzerland, Taiwan, Turkey, Uganda, UK, and Vietnam.
Recall number
Z-1528-2018
Recall status
Terminated
CDRH record ID
163904
FDA record created
2018-04-26
Recall event number
79905

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗