Z-1529-2025: DxI 9000 Access Immunoassay Analyzer, catalog number C11137 in vitro diagnostic device used for the quan

Recall record Z-1529-2025 for Beckman Coulter, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1529-2025
Company in FDA record
Beckman Coulter, Inc.
FDA classification date
Apr 4, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Ground screws on Dxl 9000 PCB boards may cause electric shorts, leading to smoke and potentially delaying patient results. Smoke also poses a hazard to equipment and users.

Source facts

Action
The firm sent notice of "Urgent Medical Device Recall" to affected consignees (all OUS), dated July 3, 2024. Customers were informed of the potential for barcode read failure error, delay of reporting patient test results, and possibility for a small amount of smoke to be emitted. Consignees were advised that no action was necessary if their laboratory has not observed any RSM Barcode Read Failure error messages. If you do observe any such error messages, or notice smoking or burning smell coming from the Dxl 9000 analyzer, please contact your Beckman Coulter service representative. Beckman Coulter will no longer ship these analyzers with the affected ground screw. Your Beckman Coulter service representative will contact you to schedule a part replacement that resolves the issue.
Initiated
2024-07-12
Root cause
Device Design
Distribution
OUS (Foreign) distribution to countries of: Australia, Croatia, Czechia, France, Germany, Ireland, Israel, Italy, Netherlands, New Zealand, South Africa, Spain, and United Kingdom of Great Britain and Northern Ireland.
Recall number
Z-1529-2025
Recall status
Open, Classified
CDRH record ID
212795
FDA classified
2025-04-04
Recall event number
96406

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗