Open original FDA record ↗Recall record
- FDA identifier
- Z-1530-2008
- Company in FDA record
- BD Biosciences
- FDA classification date
- Aug 8, 2008
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Values might not update in statistics views after certain functions are performed in a worksheet.
Source facts
- Action
- The firm issued an Important Product Information notification letter on 12/21/2007. The letter instructs users to reload data by closing and reopening an experiment before reporting results in order to determine if their statistics have been impacted. They are asked to fill out and return a form pertaining to this problem. BD Biosciences will release a version of BD FACSDiva software to correct this problem. The letter also stated that BD Biosciences rep will contact the users when the new software is available. User are asked to contact BD Customer Support Center at 1-877-232-8995 if they have any additional questions.
- Initiated
- 2007-12-21
- Root cause
- Software design
- Terminated
- 2008-09-24
- Distribution
- Nationwide Distribution.
- Recall number
- Z-1530-2008
- Recall status
- Terminated
- CDRH record ID
- 67312
- FDA classified
- 2008-08-08
- Recall event number
- 46372