Skip to content

Public FDA record

Z-1532-2025: IVX Station Reconstitution Transfer Set. Model Number: 298921215. Packaging: 1 set per White Tyvek Pouch,

Recall record Z-1532-2025 for Omnicell, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1532-2025
Company in FDA record
Omnicell, Inc.
FDA classification date
Apr 4, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

IVX Station Reconstitution Transfer Sets were shipped in advance of the IVX Station upgrade and not intended for clinical use but are engineering devices intended for use during set up and system calibration. The items were not shipped in a validated packing configuration and sterility cannot be ensured.

Source facts

Action
Single customer was notified on 02/21/2025 via email and later received the letter via hand delivery on 02/25/2025. The notification instructed the customer that Omnicell personnel will remove the affected units, and any Reconstitution Transfer Sets required for Omnicell use during set up and upgrade will be provided as part of a supply kit and will be marked not for human use. The customer was requested to complete and return the Customer Acknowledgement and Receipt.
Initiated
2025-02-20
Root cause
Process control
Distribution
Distributed in the US to Alabama
Recall number
Z-1532-2025
Recall status
Open, Classified
CDRH record ID
212828
FDA classified
2025-04-04
Recall event number
96422

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗