Open original FDA record ↗Recall record
- FDA identifier
- Z-1533-2016
- Company in FDA record
- B Braun Medical Inc
- FDA classification date
- May 3, 2016
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
Potential leakage of conductivity sensors of the Dialog Dialysis machines.
Source facts
- Action
- B. Braun mailed an Urgent Medical Device Correction notice to affected customers informing them of the Dialog+Hemodialysis System due to the potential for Bicarbonate and End conductivity cells (3456102A and 3456103A respectively) to exhibit cracks. Notice provided customers with actions for Qualified Service Technicians, Clinical Personnel Administering Dialysis treatment and adverse event reporting. All Dialog+Hemodialysis machines will be inspected at the customer's site by the customer Biomed, who has been trained and qualified by B. Braun Avitum. The inspection of each machine will be performed in accordance to a Field Service Information (FSI). Customers with questions were instructed to contact the Technical Support Team at 800-621-0445 or the Customers Service Team at 800-848-2066. For questions regarding this recall call 610-997-4717.
- Initiated
- 2016-04-01
- Root cause
- Process control
- Terminated
- 2017-07-11
- Distribution
- Nationwide Distribution to ND, OH, CA, VA, FL, LA, SC, IL, UT, AZ, KY, IA, WI, OK, TN, SD, AL, NJ, TX, NM, NY, KS, WA, MO, CO, PA, CT, DE, GA, MD, VA, Puerto Rico, and St. Thomas USVI
- Recall number
- Z-1533-2016
- Recall status
- Terminated
- CDRH record ID
- 144858
- FDA classified
- 2016-05-03
- Recall event number
- 73744