Open original FDA record ↗Recall record
- FDA identifier
- Z-1533-2025
- Company in FDA record
- ZOLL Circulation, Inc.
- FDA classification date
- Apr 11, 2025
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
AutoPulse NXT Resuscitation System may not provide adequate chest compressions or may stop compressions due to code FC1060.
Source facts
- Action
- On March 4, 2025, Zoll Circulation issued a "Medical Device Product Correction Notice" to affected consignees. Zoll asked consignees to take the following actions: 1. Stop using your affected platform(s) immediately. Your patients may not receive high-quality compressions. 2.Complete the Customer Acknowledgment/ No-Charge AP NXT Platform Repair Request Form (page 3)and email it to dlam@zoll.com within 3 days. Please feel free to call me at (408) 419 - 2955 if you have any questions. 3.Return your affected platform(s) to ZOLL as soon as you receive the loaner AP NXT Platform(s) from ZOLL.
- Initiated
- 2025-03-04
- Root cause
- Process change control
- Distribution
- US: AL, CA, CT, FL, IL, KS, LA, MN, MO, MT, NC, ND, NM, NY, OH, OK< PA, TX, WI OUS: Canada, Hong Kong,
- Recall number
- Z-1533-2025
- Recall status
- Open, Classified
- CDRH record ID
- 213013
- FDA classified
- 2025-04-11
- Recall event number
- 96479