Skip to content

Public FDA record

Z-1534-2009: TransFx External Fixation System Drill Bit, Quick Connect, 3.0 mm diameter, 180 mm length, nonsterile, Zi

Recall record Z-1534-2009 for Zimmer Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1534-2009
Company in FDA record
Zimmer Inc.
FDA classification date
Jun 30, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Drill tip is out of specifications, reducing the efficiency of drilling action, requiring more force and potentially heating the bone surface during operation.

Source facts

Action
Distributors and consignees were notified by an Urgent: Device Recall letter dated 5/1/09. Consignees were advised to stop using the device and quarantine it immediately. Distributors were instructed to carry out a physical count of all affected product and record the data on the Inventory Return Certification Form; fax a copy of the completed form to Zimmer, Inc.; and return recalled product along with the Inventory Return Certification Form to the firm.
Initiated
2009-05-01
Root cause
Process control
Terminated
2010-11-03
Distribution
Worldwide Distribution -- USA, Australia, Canada, El Salvador, England, Germany, Latin America, and Singapore.
Recall number
Z-1534-2009
Recall status
Terminated
CDRH record ID
81657
FDA classified
2009-06-30
Recall event number
51924

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗