Open original FDA record ↗Recall record
- FDA identifier
- Z-1535-2009
- Company in FDA record
- Zimmer Inc.
- FDA classification date
- Jun 30, 2009
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Drill tip is out of specifications, reducing the efficiency of drilling action, requiring more force and potentially heating the bone surface during operation.
Source facts
- Action
- Distributors and consignees were notified by an Urgent: Device Recall letter dated 5/1/09. Consignees were advised to stop using the device and quarantine it immediately. Distributors were instructed to carry out a physical count of all affected product and record the data on the Inventory Return Certification Form; fax a copy of the completed form to Zimmer, Inc.; and return recalled product along with the Inventory Return Certification Form to the firm.
- Initiated
- 2009-05-01
- Root cause
- Process control
- Terminated
- 2010-11-03
- Distribution
- Worldwide Distribution -- USA, Australia, Canada, El Salvador, England, Germany, Latin America, and Singapore.
- Recall number
- Z-1535-2009
- Recall status
- Terminated
- CDRH record ID
- 81658
- FDA classified
- 2009-06-30
- Recall event number
- 51924