Skip to content

Public FDA record

Z-1535-2015: CD99 Catalog number PM008 AA; For In Vitro Diagnostic Use. Product Usage: Provides adjunctive diagnostic

Recall record Z-1535-2015 for Biocare Medical, LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1535-2015
Company in FDA record
Biocare Medical, LLC
FDA classification date
Apr 27, 2015
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

A drop in staining intensity over time has been observed. These lots may not meet their labeled shelf life.

Source facts

Action
Biocare sent a Medical Device Notification letters dated March 17, 2015 to all affected sites. The letters identify the affected product, problem, actions to be taken and the fact that it affects all identified lots shipped between 02/13/13 and / 01/28/15. The letters instructed customers to immediately discontinue further use of the recalled lots, dispose of them according to state and local regulations and complete the response form and returned by April 7, 2015. Credit will be issued upon receipt of the response form.
Initiated
2015-03-17
Root cause
Unknown/Undetermined by firm
Terminated
2015-06-15
Distribution
Worldwide Distribution - US nationwide in the states of TX, KS, OK, NJ, LA, GA, NC, SC, VA, and the countries of Spain, Austria, Ecuador, Dominican Republic, Vietnam, Kuwait, Hungary and Israel.
Recall number
Z-1535-2015
Recall status
Terminated
CDRH record ID
134965
FDA classified
2015-04-27
Recall event number
70823

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗