Skip to content

Public FDA record

Z-1535-2017: LipiFlow Thermal Pulsation System/LipiFlow, System Activator (Disposable), Rx Only, Sterile

Recall record Z-1535-2017 for Tearscience, Inc. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1535-2017
Company in FDA record
Tearscience, Inc
FDA classification date
Mar 22, 2017
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Labeling: The labeled storage temperature is not consistent with the temperature conditions tested for the entire four year shelf life duration.

Source facts

Action
Tearscience sent an MEDICAL DEVICE LABELING CORRECTION/RECALL letter dated January 30, 2017. .Actions to be Taken by Customer: Within 10 business days of receipt of this letter, please take the following actions: 1) Read the enclosed Summary of Important Labeling Changes for the LipiFlow System. 2) Complete the Acknowledgement and Receipt form at www.TSVerify.pmcprograms.com or return the enclosed completed form by email or mail. 3) If you have stored Activators in an environment that is not temperature controlled (without heat and air conditioning), discontinue use of these Activators and call TearScience Labeling Change Hotline at (844) 872-5658. For any questions regarding this letter, please call (844) 872-5658 Monday through Friday, 8:00AM to 6:00PM, Eastern Time. We want to thank you for being a valuable TearScience partner, and as always would like to thank you for your business, your trust, and your commitment to treating MGD. For further questions please call (919) 459-4831.
Initiated
2017-01-30
Root cause
Labeling Change Control
Terminated
2017-11-27
Distribution
US (nationwide)
Recall number
Z-1535-2017
Recall status
Terminated
CDRH record ID
152977
FDA classified
2017-03-22
Recall event number
76365

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗