Skip to content

Public FDA record

Z-1536-06: Medtronic 3777 Lead Kit, 1x8 Low Impedance Lead Kit for Spinal Cord Stimulation (SCS). Manufactured at:

Recall record Z-1536-06 for Medtronic Neurological. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1536-06
Company in FDA record
Medtronic Neurological
FDA classification date
Sep 29, 2006
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Model 3777 Standard Octad Lead (Lot V009545) and Model 3778 Compact Octad Lead (Lot V009546) were mispackaged. The result of this error is that lead kits labeled 3777 lot V009545 may contain a 3778 compact-spaced lead, while those labeled 3778 lot V009546 may contain a 3777 standard-spaced lead. The result of this error is an overall increase or decrease of 14 mm in the electrode coverage.

Source facts

Action
Medtronic representatives retrieved non-implanted devices. An Urgent Medical Device Recall letter was sent to physicians to describe the issue and patient management information.
Initiated
2006-08-02
Root cause
Other
Terminated
2006-12-31
Distribution
Nationwide distribution --- including states of CA, GA, KY, LA, and TX.
Recall number
Z-1536-06
Recall status
Terminated
CDRH record ID
48106
FDA classified
2006-09-29
Recall event number
36164

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗