Post Market Surveillance of the 90 degree Cannulated Infant Blade Plate indicated that there have been five reports concerning bending of the plate intra-operatively.
Source facts
Action
OrthoPediatrics sent an Urgent Voluntary Recall notifiation via telephone and email to all affected consignees. The letter instructed consignees to take possession of any affected devices and quarantine immediately. A Mandatory Reply Form is included with the notification letter to be completed and returned. Please contact Greg Teghtmeyer if you have any questions (574-268-6379).
Initiated
2013-04-19
Root cause
Device Design
Terminated
2013-09-19
Distribution
USA Nationwide Distribution
Recall number
Z-1536-2013
Recall status
Terminated
CDRH record ID
117875
FDA classified
2013-06-12
Recall event number
65091