Open original FDA record ↗Recall record
- FDA identifier
- Z-1536-2018
- Company in FDA record
- Smith & Nephew, Inc.
- FDA record creation date
- Apr 26, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Packaging design is insufficient to prevent potential sterile barrier breach (pin hole) in Tyvek film of pouch. A limited number of previously recalled devices were inadvertently redistributed.
Source facts
- Action
- On 12/11/2017, recall notifications were sent to Hospital Representatives, Affiliates/Distributors, and Sales Personnel. Notified consignees were instructed to inspect inventory and quarantine any affected devices, return the completed Inventory Return Certification Form, and return the affected product. In addition to the above, Affiliates and Distributors were instructed to forwarding the recall letter to customers and asking them to return any recalled product. Consignees were instructed to contact the firm via email at FieldActions@smith-nephew.com
- Initiated
- 2017-12-11
- Root cause
- Employee error
- Terminated
- 2020-09-21
- Distribution
- Distributed to accounts in 7 states: AZ, ME, NY, OH, PA, TN, and VT; and PR.
- Recall number
- Z-1536-2018
- Recall status
- Terminated
- CDRH record ID
- 163060
- FDA record created
- 2018-04-26
- Recall event number
- 79700