Open original FDA record ↗Recall record
- FDA identifier
- Z-1536-2021
- Company in FDA record
- Fresenius Medical Care Holdings, Inc.
- FDA record creation date
- May 7, 2021
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
AquaBplus HF module on the AquaBplus 2500 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2500 device. The 3-ring main and fluid fly loops the heat disinfection may not be as effective as a 2-ring main installation
Source facts
- Action
- Fresenius issued Urgent Medical Device Alert (FA-2021-07-C) to customers on 03/18/2021 via First Class Mail. Letter states reason for recall, health risk and action to take: Please note that all installed AquaBplus 2500 osmosis systems are validated for chemical disinfection upon installation. You may continue to use your HF module, since the chemical disinfection remains the primary disinfection method. The system will continue to provide dialysis water that meets Association for the Advancement of Medical Instrumentation (AAMI) standards; thus, there is no risk introduced. You should continue to perform all microbiological testing per clinical policies and procedures. In the event your ring mains endotoxin or microbiological results are above action limits, please inform your Fresenius Medical Care Renal Therapies Group (FMCRTG) Water Systems Service Specialist. An FMCRTG Water Systems Service Specialist will be contacting you to schedule an evaluation of your Water Treatment system and provide feedback on your system. For questions or concerns, please contact FMCRTG Technical Support Services at 888-553-1109.
- Initiated
- 2021-03-18
- Root cause
- Labeling Change Control
- Terminated
- 2024-04-10
- Distribution
- US Nationwide distribution in the states of FL, MA, MD, MI, MO, NC, NJ, NY, PA, PR, SC, TN, TX.
- Recall number
- Z-1536-2021
- Recall status
- Terminated
- CDRH record ID
- 186699
- FDA record created
- 2021-05-07
- Recall event number
- 87659