Z-1536-2021: AquaBplus and AquaBplus B2 Water Purification System - Product Usage: Intended for use with a hemodialysi

Recall record Z-1536-2021 for Fresenius Medical Care Holdings, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1536-2021
Company in FDA record
Fresenius Medical Care Holdings, Inc.
FDA record creation date
May 7, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

AquaBplus HF module on the AquaBplus 2500 osmosis system, an error in the US Installation Guideline instructed that 3-ring mains, instead of 2-ring mains, may be installed in combination with an AquaBplus 2500 device. The 3-ring main and fluid fly loops the heat disinfection may not be as effective as a 2-ring main installation

Source facts

Action
Fresenius issued Urgent Medical Device Alert (FA-2021-07-C) to customers on 03/18/2021 via First Class Mail. Letter states reason for recall, health risk and action to take: Please note that all installed AquaBplus 2500 osmosis systems are validated for chemical disinfection upon installation. You may continue to use your HF module, since the chemical disinfection remains the primary disinfection method. The system will continue to provide dialysis water that meets Association for the Advancement of Medical Instrumentation (AAMI) standards; thus, there is no risk introduced. You should continue to perform all microbiological testing per clinical policies and procedures. In the event your ring mains endotoxin or microbiological results are above action limits, please inform your Fresenius Medical Care Renal Therapies Group (FMCRTG) Water Systems Service Specialist. An FMCRTG Water Systems Service Specialist will be contacting you to schedule an evaluation of your Water Treatment system and provide feedback on your system. For questions or concerns, please contact FMCRTG Technical Support Services at 888-553-1109.
Initiated
2021-03-18
Root cause
Labeling Change Control
Terminated
2024-04-10
Distribution
US Nationwide distribution in the states of FL, MA, MD, MI, MO, NC, NJ, NY, PA, PR, SC, TN, TX.
Recall number
Z-1536-2021
Recall status
Terminated
CDRH record ID
186699
FDA record created
2021-05-07
Recall event number
87659

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗