Open original FDA record ↗Recall record
- FDA identifier
- Z-1537-2012
- Company in FDA record
- Haemacure Corporation
- FDA classification date
- May 11, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
This product has been recalled due to the lot number being different on the box and pouch. The lot number on the box label was incorrectly printed as 11144 instead of 111144. The pouch label referenced the correct lot number, 111144.
Source facts
- Action
- Haemacure Corporation, Sarasota, FL initiated letters via e-mail on 11/16/2009 with delivery and read receipt confirmation to their two consignees. The letters identified the affected product and stated the reason for the recall. The letter also states that the firm will replace each order upon receipt of the mislabeled boxes at no charge to the customer. Customers were asked to return their unused devices. Questions should be directed to customer service at 1-877-872-4583.
- Initiated
- 2009-11-16
- Root cause
- Labeling mix-ups
- Terminated
- 2012-05-14
- Distribution
- Nationwide Distribution -- Illinois and California.
- Recall number
- Z-1537-2012
- Recall status
- Terminated
- CDRH record ID
- 108601
- FDA classified
- 2012-05-11
- Recall event number
- 61636