Open original FDA record ↗Recall record
- FDA identifier
- Z-1538-2012
- Company in FDA record
- Philips Healthcare Inc.
- FDA classification date
- Jun 13, 2012
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 17, 2026
- Recall classification
- Class II
Reason stated in the source
When the system is switched off or when the tube arm is parked in the down position (X-ray generator switch off), a radiation pulse (50 kV, 2.5 pGy) is generated inadvertently. .When the system is switched off or when the tube arm is parked in the down position (X-ray generator switch off), a radiation pulse (50 kV, 2.5 pGy) is generated inadvertently. The failure happened when component supplie
Source facts
- Action
- Philips Healthcare sent a proprosed corrective action letter dated May 3, 2012, to FDA. The proposed corrective action plan is to bring the product into compliance with regulatory requirements by issuing a field corrective action which should be completed by Philips by August 1, 2012. For questions regarding this recall call 978-687-1501.
- Initiated
- 2011-10-19
- Root cause
- Radiation Control for Health and Safety Act
- Terminated
- 2017-01-18
- Distribution
- Nationwide Distribution
- Recall number
- Z-1538-2012
- Recall status
- Terminated
- CDRH record ID
- 109197
- FDA classified
- 2012-06-13
- Recall event number
- 61830