Skip to content

Public FDA record

Z-1539-05: XSYSTEMS AxSYM Multiconstituent Controls for Abused Drug Assays, list 9687-12; Abbott laboratories, Abbot

Recall record Z-1539-05 for Abbott Laboratories MPG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1539-05
Company in FDA record
Abbott Laboratories MPG
FDA classification date
Sep 14, 2005
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The XSYSTEMS Multiconstituent Controls accessory insert (commodity number 34-3067/R6) does not contain the correct control ranges for the AxSYM Opiates assay.

Source facts

Action
On 6/27/05, Abbott sent the customer letter via FedEx to all XSYSTEMS Multiconstituent Controls customers that received List Number 9687-12 lots 17270M100 and/or 18177M300 and/or AxSYM Opiates reagents (List Number 3B25-20) since June 01, 2004. The letters explained that the multiconstituent control ranges for morphine were not updated in the MCC Accesory insert at the same time they were updated in the AxSYM Opiates reagent package insert in 2004. The accounts were requested to review their current control ranges for the AxSYM Opiates assay, and verify that they are using either the control ranges in the AxSYM Opiates reagent package insert or control ranges established by their facility. They were also instructed to follow their facility''s normal policies and procedures for communicating with health care providers served by their laboratory. The accounts were requested to complete and return the enclosed reply sheet acknowledging the receipt of the letter and the dissemination of the letter to the laboratory manager/supervisor responsible for the assays.
Initiated
2005-06-27
Root cause
Other
Terminated
2006-01-18
Distribution
Nationwide, including Puerto Rico, and internationally to Argentina, Australia, Barbados, Bermuda, Brazil, Canada, Cayman Island, Chile, China/Hong Kong, Columbia, Costa Rica, Dominican Republic, Ecuador, Guatemala, Germany, Honduras, Jamaica, Korea, Mexico, New Zealand, Panama, Peru, Singapore, St. Vincent, Thailand, Trinidad and Taiwan
Recall number
Z-1539-05
Recall status
Terminated
CDRH record ID
40299
FDA classified
2005-09-14
Recall event number
32540

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗