Skip to content

Public FDA record

Z-1539-2014: Infant Nasal CPAP (continuous positive airway pressure) Prong, Nasal. CPAP Prongs (10 pack): Model No. B

Recall record Z-1539-2014 for Fisher & Paykel Healthcare, Ltd.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1539-2014
Company in FDA record
Fisher & Paykel Healthcare, Ltd.
FDA classification date
May 20, 2014
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

The Infant Nasal CPAP Prong has the potential to detach from the Nasal Tubing and therapy is likely to be interrupted.

Source facts

Action
Fisher & Paykel Healthcare sent an Urgent Medical Device Recall letter on April 11, 2014, to all affected customers. The letter informed them of the problems identified and the actions to be taken. Customers were instructed to complete Section A 'Inspection of Stock' on the Urgent Medical Device Recall Response Form and return it to their FPH Representative. Destroy the Affected Prongs by cutting them in half. Customers were also instructed to contact their FRH Representative for replacement product. For questions regarding this recall call 949-453-4000, ext 1423.
Initiated
2014-04-15
Root cause
Device Design
Terminated
2014-09-30
Distribution
Worldwide Distribution - USA (nationwide) and Internationally to: ARGENTINA AUSTRALIA BANGLADESH BELGIUM CANADA CHILE CHINA FRANCE GERMANY GREAT BRITAIN GREECE HUNGARY INDIA INDONESIA IRAQ ISRAEL ITALY JAPAN LIBYA MALAYSIA NEPAL NETHERLANDS NEW ZEALAND PERU POLAND PORTUGAL QATAR ROMANIA RUSSIA RWANDA SAUDI ARABIA SOUTH AFRICA SPAIN SRI LANKA SWITZERLAND TAIWAN THAILAND TRINIDAD & TOBAGO TURKEY UNITED ARAB EMIRATES
Recall number
Z-1539-2014
Recall status
Terminated
CDRH record ID
126776
FDA classified
2014-05-20
Recall event number
68011

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗