Open original FDA record ↗Recall record
- FDA identifier
- Z-1539-2014
- Company in FDA record
- Fisher & Paykel Healthcare, Ltd.
- FDA classification date
- May 20, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class I
Reason stated in the source
The Infant Nasal CPAP Prong has the potential to detach from the Nasal Tubing and therapy is likely to be interrupted.
Source facts
- Action
- Fisher & Paykel Healthcare sent an Urgent Medical Device Recall letter on April 11, 2014, to all affected customers. The letter informed them of the problems identified and the actions to be taken. Customers were instructed to complete Section A 'Inspection of Stock' on the Urgent Medical Device Recall Response Form and return it to their FPH Representative. Destroy the Affected Prongs by cutting them in half. Customers were also instructed to contact their FRH Representative for replacement product. For questions regarding this recall call 949-453-4000, ext 1423.
- Initiated
- 2014-04-15
- Root cause
- Device Design
- Terminated
- 2014-09-30
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to: ARGENTINA AUSTRALIA BANGLADESH BELGIUM CANADA CHILE CHINA FRANCE GERMANY GREAT BRITAIN GREECE HUNGARY INDIA INDONESIA IRAQ ISRAEL ITALY JAPAN LIBYA MALAYSIA NEPAL NETHERLANDS NEW ZEALAND PERU POLAND PORTUGAL QATAR ROMANIA RUSSIA RWANDA SAUDI ARABIA SOUTH AFRICA SPAIN SRI LANKA SWITZERLAND TAIWAN THAILAND TRINIDAD & TOBAGO TURKEY UNITED ARAB EMIRATES
- Recall number
- Z-1539-2014
- Recall status
- Terminated
- CDRH record ID
- 126776
- FDA classified
- 2014-05-20
- Recall event number
- 68011