Open original FDA record ↗Recall record
- FDA identifier
- Z-1539-2017
- Company in FDA record
- Integra LifeSciences Corp.
- FDA record creation date
- Mar 23, 2017
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Voluntary recall/Removal of Panta and Panta XI Nails because there may be the potential that the internal and external package seal does not provide the level of sterility intended.
Source facts
- Action
- Integra sent "Urgent Medical Device Recall Letters and Return Response Forms" dated March 3, 2017 to their distributors and customers. All customers were asked to examine their inventory to determine if they had the recalled lots and if they did they are instructed to stop using them immediately. Distributors are asked if the recalled product was further distributed, those customers should be notified and instructed to follow the recalling procedures. The return response form should be completed, even if they do not have any recalled lots, annotating the amount of product and list a record of the lot numbers. The return response form should be sent back to Integra via email to: FCA2@integralife.com or via fax to: 1-609-750-4220. Upon receipt by Integra, if your company has identified recalled lots, Customer Service will contact the customer and provide an RMA number and directions to return the product, as well as input an order to replace the quantity indicated on the return response form. Should any customer have any questions regarding the recall instructions, they can contact Customer Service at 1-800-654-2873. To return units under field inventory, customer can contact 1-888-601-0203, option 2. For further questions, please call (717) 309-3966.
- Initiated
- 2017-03-03
- Root cause
- Packaging change control
- Terminated
- 2018-05-21
- Distribution
- US Nationwide and Internationally
- Recall number
- Z-1539-2017
- Recall status
- Terminated
- CDRH record ID
- 153928
- FDA record created
- 2017-03-23
- Recall event number
- 76693