Open original FDA record ↗Recall record
- FDA identifier
- Z-1540-2014
- Company in FDA record
- Alere San Diego, Inc.
- FDA classification date
- May 2, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
These lots may not recover within range when tested using certain commercially available controls.
Source facts
- Action
- Alere San Diego, Inc. initiated this recall by sending out notification letters via email and/or regular mail on March 20, 2014. The letter titled "URGENT MEDICAL DEVICE CORRECTION", dated "March 20, 2014, informed customers of the reason for recall, and potential impact, product description with lots numbers, customer/distributor required action, and contact information. The letter was accompanied by a "Customer/Distributor Verification Form". Customers were instructed that they may continue to use reagent lots 329599 and 331265. " If they are using an alternate commercial control with Reagent Pack Lots 329599 or 331265, they can receive the Alere Triage BNP QC Controls PN 98201 by requesting them on the Verification Form. "Customers were instructed to share this information with their laboratory staff, including their laboratorys medical director and retain this notification as part of their laboratory Quality System documentation. For questions regarding this recall call 858-805-2000, ext 3015. " If they have forwarded the product listed above to another laboratory, please provide a copy of this letter to them. " Complete and FAX the enclosed Verification Form within 10 days to confirm your receipt of this notice. Only one Verification Form is required per facility. Customers with questions were instructed to call 877-308-8287 or email Triage.Support@alere.com.
- Initiated
- 2014-03-20
- Root cause
- Under Investigation by firm
- Terminated
- 2015-01-23
- Distribution
- Worldwide Distribution - USA (nationwide) and Internationally to Australia, Canada, China, EU, Mexico, Russia, Singapore, South Africa, South Korea, Taiwan and United Arab Emirates.
- Recall number
- Z-1540-2014
- Recall status
- Terminated
- CDRH record ID
- 126592
- FDA classified
- 2014-05-02
- Recall event number
- 67910