Z-1541-2009: i-STAT G3+ Cartridge; pH, PCO, PO. HCO2, TCO2, BE sO2; Test includes: PCO2 for calculated HCO3 and PO2 fo

Recall record Z-1541-2009 for Abbott Point of Care Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1541-2009
Company in FDA record
Abbott Point of Care Inc.
FDA classification date
Jun 18, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Equipment problem: i-STAT cartridges were not properly sealed as the packaging system jammed, causing compromised seals.

Source facts

Action
Urgent Recall Notice letters pertaining to the EG7+ Cartridges were sent on February 23, 2009 by first class mail to all customers. The firm found one other affected lot of G3+ cartridges and expanded the recall to include one other lot. Letters for the expansion are dated May 2009. Customers were asked to use the attached form to record all of the box numbers if they have any cartridges from the affected lot. Cartridges from the box numbers listed will be replaced an any remaining unused cartridges from those specific boxes are to be returned as instructed on the form. To receive replacement cartridges, customers are to follow the instructions on the enclosed customer reply form. If any cartridges from the affected lot were forwarded to another facility, a copy of the recall letter needs to be provided to them and the information at the bottom of the customer reply form should be completed. Questions should be directed to 1-800-366-8020, Option 1.
Initiated
2009-02-23
Root cause
Packaging process control
Terminated
2009-07-29
Distribution
Worldwide Distribution -- USA, Austria, China, India, Saudi Arabia, South Africa, and Taiwan.
Recall number
Z-1541-2009
Recall status
Terminated
CDRH record ID
81617
FDA classified
2009-06-18
Recall event number
51211

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗