- FDA identifier
- Z-1541-2010
- Company in FDA record
- Encision, Inc.
- FDA classification date
- May 6, 2010
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 20, 2026
- Recall classification
- Unknown — no matched enforcement classification
Reason stated in the source
Electrosurgical electrode tip insulation may crack and break off during use.
Source facts
- Action
- A "Confirmation of Product Recall" letter dated April 5, 2010, was faxed by the firm (Encision, Inc.) to all customers. The letter describes the product, problem and action to be taken by customers. The customers were instructed to either return or destroy the affected electrodes and to contact any customers or sales representatives to whom they distributed the product and confirm they have been removed from use and/or they will pick them up for return. The customers is to complete form and fax to the attention of RA/QA at 1-303-444-2693 or return with the items. The number to call to arrange return and/or if you have any questions is 1-800-998-0986.
- Initiated
- 2010-04-05
- Root cause
- Nonconforming Material/Component
- Terminated
- 2010-05-21
- Distribution
- Worldwide distribution: USA including states of CA, CT, IA, IN, KS, KY, MI, OK, TX, WV, and countries of New Zealand and The Netherlands.
- Recall number
- Z-1541-2010
- Recall status
- Terminated
- CDRH record ID
- 90481
- FDA classified
- 2010-05-06
- Recall event number
- 55286