Z-1541-2018: Specimen Gate Screening Center, Part Number 5002-0500, All software versions from 1.0 to 1.8 (current sof

Recall record Z-1541-2018 for PerkinElmer Life and Analytical Sciences, Wallac, OY. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1541-2018
Company in FDA record
PerkinElmer Life and Analytical Sciences, Wallac, OY
FDA record creation date
Apr 27, 2018
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Potential errors in patient results generated by the Screening Center product that include both false negative and false positive results.

Source facts

Action
PerkinElmer sent an Urgent Medical Device Correction letter dated November 14,2017. Instructions: 1. Implement additional control measures concerning demographic information, by taking one of the following actions a. Fully enter demographics before loading specimens on instruments (i.e. specimens can be punched, but the assay plates cannot be loaded to the measurement instruments). b. If the above mentioned is not possible or if previously entered demographic information changes after loading the assay plates on the measurement instruments, recalculate all your results in Specimen Gate laboratory Result Viewer prior to assay acceptance. 2. Follow unreported specimens with available tools to avoid any delay in reporting screening results. For further questions, please call (317) 418-1735
Initiated
2017-11-17
Root cause
Software design
Terminated
2021-08-12
Distribution
Worldwide Distribution - US Distribution to the state of Georgia., and to the countries of : Belgium, Canada, Denmark, Italy, and United Kingdom.
Recall number
Z-1541-2018
Recall status
Terminated
CDRH record ID
162274
FDA record created
2018-04-27
Recall event number
79421

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗