Open original FDA record ↗Recall record
- FDA identifier
- Z-1542-2015
- Company in FDA record
- Spacelabs Healthcare Inc
- FDA classification date
- Apr 28, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Ultraview SL Command Modules which were upgraded with the Masimo SpO2 Option, Model 91496-M was affected. Customer reported that during the time that the module is resetting (~10 seconds), monitoring of all parameters will be suspended. Upon completion of this sequence, alarms will be reset to their default values.
Source facts
- Action
- Spacelabs Healthcare sent an Urgent - Medical Device Correction letter, dated 17 April 2015, to the US consignees via priority mail. On 4/24/2015, the firm will send the Urgent - Medical Device Correction letter (translated as necessary) via email to international subsidiaries and distributor of record. The Identified the affected product, problem and actions to be taken. Ultraview SL Command Modules which were upgraded with the Masimo SpO2 Option, Model 91496-M are affected. Ultraview SL Command Modules originally shipped with Masimo SpO2 are NOT affected. Spacelabs Healthcare will contact you to schedule a convenient time for Spacelabs to correct all of your facility's affected Command Modules at no cost. For additional information or technical assistance, please contact: Technical Support Spacelabs Healthcare, Inc. 35301 SE Center St. Snoqualmie, WA 98065 1-800-522-7025 and select 2 for Technical Support
- Initiated
- 2015-04-17
- Root cause
- Software design
- Terminated
- 2015-11-13
- Distribution
- Worldwide Distribution- US Nationwide in the states of AL, DE, GA, ID, IL, KY, LA, MA, ME, MI, MN, MO, NC, NJ, NY, OH, OR, PA, SC, TN, VA, WA and the country of France.
- Recall number
- Z-1542-2015
- Recall status
- Terminated
- CDRH record ID
- 136025
- FDA classified
- 2015-04-28
- Recall event number
- 71074