Skip to content

Public FDA record

Z-1542-2021: Regard Dressing Change Tray

Recall record Z-1542-2021 for ROi CPS LLC. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1542-2021
Company in FDA record
ROi CPS LLC
FDA record creation date
May 7, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

ChloraPrep applicators in the kit can grow organisms that can result in a breach of the applicators package integrity.

Source facts

Action
On March 26, 2021 the firm sent a notification to customers: Actions to be taken by customers: 1. CHECK: please review your inventory to determine if the manufacturer item numbers and lot numbers listed in the above table are in your inventory. 2. ACT: If the affected kits are identified, immediately quarantine. Use the attached Avery label template to label the affected kits. Place the label in a location where it is clearly visible. The label instructs the end user to remove the ChloraPrep from the kit and pull from sterile supply. 3. REPLY: Please complete the attached reply form indicating any inventory on hand. If you do not have inventory, please return the form indicating 0 affected inventory at your facility. It is important that we promptly receive your completed reply form. 4. NOTIFY: This notice is to be sent to end user facilities who were shipped the affected kits. Please immediately notify any customers to whom you have distributed or forwarded product affected by this medical device correction notice. You may include copy of this notice and the attached label template with your customer notification. 5. END USERS: Notices sent to end users should instruct that the ChloraPrep contained within the kits be removed and discarded.
Initiated
2021-03-26
Root cause
Package design/selection
Terminated
2023-01-30
Distribution
US Nationwide distribution in the state of Louisiana.
Recall number
Z-1542-2021
Recall status
Terminated
CDRH record ID
186693
FDA record created
2021-05-07
Recall event number
87658

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.