Open original FDA record ↗Recall record
- FDA identifier
- Z-1543-2015
- Company in FDA record
- Nihon Kohden America Inc
- FDA classification date
- Apr 29, 2015
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Nihon Kohden America (NKA) is recalling the Remote Network Station (RNS) 9703 because it may fail to sound.
Source facts
- Action
- The firm, Nihon Kohden, sent an Urgent Medical Device Alert" letter dated 4/14 /15 to customers. The letter describes the product, problem and actions to be taken. The customers were instructed to stop using the RNS for primary alarm notification and instead utilize the CNS and or the Bedside monitor for primary alarm notification, until such time NKA can provide corrective measures to your facility; Inform potential users in your organization of this notification, and complete the attached reply form and return it by fax (949) 580-1550 or email it to RNS_Replay@nkusa.com. Customers with questions are instructed to contact NKA Technical Support Center at (800) 325-0283 Ext 6 (available 24 hours a day/7 days/week).
- Initiated
- 2015-04-14
- Root cause
- Under Investigation by firm
- Terminated
- 2015-10-01
- Distribution
- US Distribution to states of: AL, CO, DE, FL, GA, HI, KY, IL, IN, MA, MD, MI, MN, MO, NC,, ND, NJ, NY.OH, PA, and WV.
- Recall number
- Z-1543-2015
- Recall status
- Terminated
- CDRH record ID
- 135847
- FDA classified
- 2015-04-29
- Recall event number
- 71015