Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.
Source facts
Action
On 02/26/2021, the one customer was notified of the recall via phone call and email. On 04/15/21, a Voluntary Field Safety Corrective Action notice were mailed. The firm asked their customer to respond so the bezel repair part could be replaced. Customers can contact the recalling firm at: 657-267-0054 or stdoan@stepharmedical.com
Initiated
2021-02-26
Root cause
Component design/selection
Terminated
2024-08-15
Distribution
IA
Recall number
Z-1543-2021
Recall status
Terminated
CDRH record ID
186977
FDA classified
2021-05-10
Recall event number
87773