Z-1543-2025: Stimulation RC Clinician Programmer Application, Model A71400, when used with the Inceptiv (Model 977119)

Recall record Z-1543-2025 for Medtronic Neuromodulation. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1543-2025
Company in FDA record
Medtronic Neuromodulation
FDA classification date
Apr 8, 2025
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

There is a software issue that can permanently disable communication with an implantable neurostimulator (INS) when a specific intraoperative programming sequence is used.

Source facts

Action
An "URGENT: MEDICAL DEVICE CORRECTION" notification letter dated March 2025 was sent to customers. The letter described the product, problem and actions to be taken. Required Actions: When utilizing any of the lead options mentioned above, avoid using the Switch Device feature to transfer data from the Model A71300 Stimulation Trialing CP App to the Model A71400 Stimulation RC CP App. As an alternative, manually enter the patient and lead information using the Model A71400 Stimulation RC CP App, ensuring not to utilize the Switch Device feature. Inform relevant medical staff about the potential of this issue and the steps to mitigate it. Ensure that backup INS devices are readily available for scheduled surgeries. During the implant procedure, check the system integrity before securing the neurostimulator in place as described in the Inceptiv/Intellis Pro Implant Manual. Please complete and return the customer confirmation form enclosed in this letter acknowledging that you have received this information. Please contact Medtronic Technical Services at 1-800-707-0933 if you have any questions regarding this communication.
Initiated
2025-03-04
Root cause
Under Investigation by firm
Distribution
Worldwide Distribution: US (Nationwide) and OUS (Global) to countries of: Austria, Belgium, Bulgaria, Canary Islands, Cyprus, Croatia, Czech Republic, Denmark, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Japan, Luxembourg, Malta, Netherlands, Northern Ireland, Norway, Poland, Portugal, Puerto Rico, Slovakia, Slovenia, Spain, Sweden, Switzerland, Turkey, and United Kingdom.
Recall number
Z-1543-2025
Recall status
Open, Classified
CDRH record ID
212886
FDA classified
2025-04-08
Recall event number
96445

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗