Skip to content

Public FDA record

Z-1544-2009: BrainLab Micro-Multileaf-Collimator, model 41600; BrainLab AG, Kapellenstrasse 12, 85622 Feldkirchen, Ger

Recall record Z-1544-2009 for Brainlab AG. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1544-2009
Company in FDA record
Brainlab AG
FDA classification date
Jul 2, 2009
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

The Integrated Multileaf-Collimator (MLC) of the linear accelerator could interfere with the intended conformal fields of the BrainLab m3 micro-Multileaf-Collimator (mMLC) during patient treatment.

Source facts

Action
BrainLAB AG issued an "Urgent Field Safety Notice/Product Notification" letter dated March 19, 2009 informing all customers of the affected device. Consignees were instructed on proper steps to take to avoid further issues. In addition, the most current m3 mMLC user guide, version 7.4, was scheduled to be sent to users in April 2009. For further questions, contact BrainLAB AG by phone at 1-800-597-5911).
Initiated
2009-03-19
Root cause
Other
Terminated
2010-08-19
Distribution
Nationwide Distribution - AL, CA, CO, FL, GA, IL, KS, KY, LA, MD, MI, MN, MS, ME, NJ, NY, OR, TN, TX, UT, VA, WV, WA and WI.
Recall number
Z-1544-2009
Recall status
Terminated
CDRH record ID
80516
FDA classified
2009-07-02
Recall event number
51484

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗