Open original FDA record ↗Recall record
- FDA identifier
- Z-1544-2014
- Company in FDA record
- Stryker Endoscopy
- FDA classification date
- May 2, 2014
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Gravity sterilization parameters provided in the 12 Instrument Sterilization reprocessing guide did not fully consider all worst-case scenarios during validation testing.
Source facts
- Action
- Stryker sent an Urgent Device Corrections letters dated April 18, 2014 to all US customers via certified mail. Stryker Sales representative and Stryker international distributors were notified with the letter and response card by e-mail. International sites are responsible for notifying customers. The letter identified the affected product, problem, potential hazards and actions to be taken. Customers were instructed to complete and sign the enclosed Acknowledgement of Receipt form and scan and e-mail to endorecall@stryker.com. For questions call 1-800-624-4422 option 3 or via email at endorecall@stryker.com.
- Initiated
- 2014-04-16
- Root cause
- Reprocessing Controls
- Terminated
- 2014-08-18
- Distribution
- Worldwide distribution
- Recall number
- Z-1544-2014
- Recall status
- Terminated
- CDRH record ID
- 126854
- FDA classified
- 2014-05-02
- Recall event number
- 68045