Skip to content

Public FDA record

Z-1544-2021: Alaris Pump Module, Model 8100, serviced/repaired by Infusion Pump Repair with an impacted bezel repair p

Recall record Z-1544-2021 for Infusion Pump Repair. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1544-2021
Company in FDA record
Infusion Pump Repair
FDA record creation date
May 10, 2021
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class I

Reason stated in the source

Bezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.

Source facts

Action
On 03/26/2021, Voluntary Field Safety Corrective Action notices started being mailed to customers. The firm conducted further communications via email and phone. The firm asked customers with pumps serviced with the affected bezels to return the pump module units to the recalling firm so they could swap out the affected bezels or provide a refund. The firm asked customers to immediately share this field safety corrective action notification with all necessary parties within your organization to ensure that they are also aware of this action. Also, please complete and return the Customer Response Form. Customers can contact the recalling firm at: 855-477-8866 or info@infusionpumprepair.com
Initiated
2021-03-26
Root cause
Component design/selection
Distribution
California
Recall number
Z-1544-2021
Recall status
Open, Classified
CDRH record ID
186689
FDA record created
2021-05-10
Recall event number
87657

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗