Z-1544-2026: PHYSIO CONTROL INFANT/CHILD Reduced Energy Defibrillator Electrodes Catalog Numbers: 11101-000016 UDI-DI

Recall record Z-1544-2026 for Physio-Control, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1544-2026
Company in FDA record
Physio-Control, Inc.
FDA classification date
Mar 10, 2026
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 16, 2026
Recall classification
Class II

Reason stated in the source

Due to pediatric defibrillator electrode delamination

Source facts

Action
On 01/21/2026, the firm sent via certified mail an "Urgent: Medical Device Recall" letter to customer informing them that Stryker was notified of complaints that the pediatric electrodes were experiencing gel delamination and could lead to a noisy or distorted ECG waveform and in addition may deliver energy to a patient without the electrode's gel barrier may result in burns. Customers are instructed to: 1. Immediately check internal inventory to locate the affected product as listed above. 2. Stop use and segregate and quarantine the affected product. 3. On the business reply form, indicate if an affected electrode is in their possession. 4. Return the business reply form by email to RSRecall@Stryker.com to confirm receipt of this notification. 5. Upon receipt of the completed business reply form, Stryker will contact you to arrange for replacement of the electrode. 6. A return material authorization (RMA) will be provided when a replacement electrode is sent to return the affected electrodes back to Stryker. 7. Maintain awareness of this communication internally until all required actions have been completed within their facility. Ensure this letter is kept with the affected device until the correction has been completed. 8. Inform Stryker if any of the subject electrodes have been distributed to other organizations. If so, provide contact details, so Stryker can inform the recipients appropriately. For Questions - contact Customer Service +1 800 787 9537, option 2, from 8:00 AM to 7:00 PM (Eastern Time), Monday Friday or by email at medtechsup@stryker.com.
Initiated
2026-01-21
Root cause
Unknown/Undetermined by firm
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AK, AL, AZ, CA, CO, CT, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, and WV. The countries of Australia, Canada, Hong Kong, Japan, Mexico, Malaysia, Netherlands, New Zealand, and Taiwan.
Recall number
Z-1544-2026
Recall status
Open, Classified
CDRH record ID
218033
FDA classified
2026-03-10
Recall event number
98257

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗