Captiva Spine has initiated a correction of the SmartLOX Instructions for Use (IFU) for SmartLOX Cervical Plate System. The pre vacuum sterilization cycle [270 (132 C) for 3 minutes] has not been validated.
Source facts
Action
Captiva Spine sent a "VOLUNTARY CORRECTION NOTIFICATION" letter dated March 23, 2012 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter provides steps for the correction and a verification form for customers to complete. Customers should contact the firm at 877-772-5571 for questions concerning this notice.
Initiated
2012-03-22
Root cause
Labeling False and Misleading
Terminated
2013-11-18
Distribution
Nationwide Distribution including the states of FL, IN, UT, GA, PA, TX, AL, NC, MI, and MN.
Recall number
Z-1545-2012
Recall status
Terminated
CDRH record ID
109124
FDA classified
2012-05-14
Recall event number
61804