Open original FDA record ↗Recall record
- FDA identifier
- Z-1545-2018
- Company in FDA record
- Becton Dickinson & Co.
- FDA record creation date
- Apr 28, 2018
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class III
Reason stated in the source
The package insert incorrectly lists the volume of the primer and probe diluent to use for the reconstitution of the master mix tube. Adding the incorrect amount of diluent could have an effect on the assay.
Source facts
- Action
- On March 2, 2018, the firm sent Urgent Medical Device Product Advisory notices to customers via UPS. The notice identified the errors and the correct information, as well as potential health risks. Customers were asked to take the following actions: 1) Share the information with all users of the product within their facility to ensure awareness; 2) Complete the customer response form and return to the recalling firm; and 3) Report any associated adverse health consequences to the recalling firm. The recalling firm is not asking customer to send back product or discard. The recalling firm will provide users with an updated insert when available. BD Customer/Technical Support: 800-638-8663, Monday-Friday, 8:30AM-5PM (EST) for US customers.
- Initiated
- 2018-03-02
- Root cause
- Error in labeling
- Terminated
- 2018-07-02
- Distribution
- Distributed to accounts in 8 states: CA, FL, NJ, NY, OR, PA, VA, and WI. International distribution to Argentina, Australia, Brazil, Canada, Chile, China, Colombia, Europe, Hong Kong, Japan, Singapore, Taiwan, and Thailand.
- Recall number
- Z-1545-2018
- Recall status
- Terminated
- CDRH record ID
- 163348
- FDA record created
- 2018-04-28
- Recall event number
- 79769