Open original FDA record ↗Recall record
- FDA identifier
- Z-1545-2020
- Company in FDA record
- Medline Industries Inc
- FDA record creation date
- Mar 20, 2020
- First observed by MedTech Watch
- Nov 16, 2026
- Latest observed content change
- Nov 16, 2026
- Recall classification
- Class II
Reason stated in the source
Voids were discovered in the packaging seal and loss of sterility on the outside of the contents within the Tyvek Form-Fill-Seal packaging.
Source facts
- Action
- Medline notified customers on February 18, 2020 via "Urgent Recall Immediate Action required" letter. The consignees were instructed to complete the response form and return the product. The recall letter identified the affected product and lot numbers. Reason for the correction was also noted in the recall letter.
- Initiated
- 2020-02-10
- Root cause
- Packaging
- Terminated
- 2023-02-13
- Distribution
- US Nationwide: MA, LA, IL, MO, MI, MN, NJ, KS, GA, NY, PA, SD, CT, TN, MD, OH, VA, FL, TX, KY, CO, IN, AL, CA, MT, OR, AK, AL, DC, IA, ME, NV, NE, SC, WV, WY, RI, NC, AZ, UT, DE and WI.
- Recall number
- Z-1545-2020
- Recall status
- Terminated
- CDRH record ID
- 180039
- FDA record created
- 2020-03-20
- Recall event number
- 85022