Z-1546-2008: Selenia Full Field Digital Mammography System with Software v. 3.3.1.1; Full field digital, system, x-ray

Recall record Z-1546-2008 for Lorad, A Hologic, Inc.. View source facts, labeled dates, and the original FDA record.

FDA identifier
Z-1546-2008
Company in FDA record
Lorad, A Hologic, Inc.
FDA classification date
May 28, 2008
First observed by MedTech Watch
Nov 16, 2026
Latest observed content change
Nov 20, 2026
Recall classification
Unknown — no matched enforcement classification

Reason stated in the source

Marker Misalignment: The new software version contained a magnification factor that shows CAD markers misaligned with the identified indications. The marked area is misaligned up to 2cm from the indication.

Source facts

Action
Consignees were notified by an Urgent Recall Notice sent on 3/21/08. The letter informed users that representatives were dispatched to all locations and they have corrected the magnification factor problem. The letter instructed users to re-evaluate any cases read using CAD from the period of installation until the date of system correction. The letter also instructed users that only images read with CAD markers are affected. The letter requests users to return a confirmation of notification. For additional information, contact 1-877-371-4372.
Initiated
2008-02-28
Root cause
Software design
Terminated
2009-02-03
Distribution
Worldwide Distribution including USA and countries of Canada and the Republic of Korea.
Recall number
Z-1546-2008
Recall status
Terminated
CDRH record ID
69385
FDA classified
2008-05-28
Recall event number
47386

Source facts and excerpts are reproduced for research. Dates retain their source meanings. “First observed” describes collection, not FDA publication or product launch.

Recall classification is supplied only when the recall number matches an enforcement record. Recall class and device class are different fields. Several recall records may belong to one recall event.

Useful context

Device classifications

FDA dataset:

Last successful check: Nov 22, 2026

Source classifications may change.

510(k) clearances

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset monthly.

Recall records

FDA dataset:

Last successful check: Nov 22, 2026

FDA publishes this dataset weekly.

Keep this research moving.

Follow ten competitors and ten product codes. Get matching changes in your dashboard and a weekly brief, for $49/month.

Start a 14-day free trial ↗